Improved encoding pulses for Bloch-Siegert B1(+) mapping.

Jankiewicz M, Gore JC, Grissom WA
J Magn Reson. 2013 226: 79-87

PMID: 23220183 · PMCID: PMC3561618 · DOI:10.1016/j.jmr.2012.11.004

A new family of optimized encoding pulses for Bloch-Siegert (BS)

B(1)(+)

mapping is introduced, as well as an algorithm to design them. The pulses are designed by numerical maximization of BS sequence sensitivity, subject to constraints that ensure low on-resonance excitation. The pulses are in all cases characterized by a constant envelope and U-shaped frequency sweep. They are validated in simulations, 7T in vivo experiments, and an experiment to measure their on-resonance excitation, and are compared to a Fermi pulse conventionally used in the BS method. The pulses are shown to produce larger phase shifts in a shorter time and with lower on-resonance excitation than the Fermi pulse, which results in lower SAR and improved

B(1)(+)

accuracy in areas of the body with large main field inhomogeneities.

Copyright © 2012 Elsevier Inc. All rights reserved.

MeSH Terms (9)

Algorithms Brain Mapping Computer Simulation Electromagnetic Fields Humans Image Enhancement Magnetic Resonance Imaging Phantoms, Imaging Reproducibility of Results