Karen Joos
Last active: 4/3/2019

Miniature real-time intraoperative forward-imaging optical coherence tomography probe.

Joos KM, Shen JH
Biomed Opt Express. 2013 4 (8): 1342-50

PMID: 24009997 · PMCID: PMC3756577 · DOI:10.1364/BOE.4.001342

Optical coherence tomography (OCT) has a tremendous global impact upon the ability to diagnose, treat, and monitor eye diseases. A miniature 25-gauge forward-imaging OCT probe with a disposable tip was developed for real-time intraoperative ocular imaging of posterior pole and peripheral structures to improve vitreoretinal surgery. The scanning range was 2 mm when the probe tip was held 3-4 mm from the tissue surface. The axial resolution was 4-6 µm and the lateral resolution was 25-35 µm. The probe was used to image cellophane tape and multiple ocular structures.

MeSH Terms (0)

Connections (1)

This publication is referenced by other Labnodes entities:

Links